Dynamic Image Analyzer for Wet Measurements Measurement: Particle shape and size Particle size range: 4 - 400µm Technology: Dynamic Image Analysis

01 มิถุนายน 2566

ผู้ชม 150 ผู้ชม

BeVision W1

BeVision W1 is a dynamic image particle size and shape analyzer for wet measurements. Based on sheath flow theory, the BeVision W1 particle image analyzer can capture each particle and deliver accurate image data. BeVision W1 particle analyzer is the best solution for scientific research and quality control. Based on a high-speed camera and sheath flow design, the BeVision W1 captures a clear image of each particle, delivering accurate particle size analysis and particle shape measurement results.

Features and Benefits

          ● Measurement range: 4 to 400 µm

          ● 24 different particle size and shape parameters

          ● Dispersion type: Wet

          ● Sheath flow technology generates high-quality image

          ● Results in compliance with ISO 9276-6

          ● Powerful software eases your work

          ● Customizable reports for different evaluation options

Features

1) Analysis parameters:
Particle size distribution, D100, aspect ratio, circularity and radius-thickness ratio.

2) Sheath flow theory: 
Ensures each individual particle passes through the focal plane of the cell sequentially, eliminates particle overlapping and defocus issues.

BeVision W1 Sheath flow theory

3) Optimized software: 
Identify
10,000 particles per minute. Automatic recognition of agglomerated powder further improves accuracy.

BeVision W1
Size Range 4-400 µm
Repeatability error ≤1% (CRM D50)
Accuracy error ≤1% (CRM D50)
Magnifying times 150-1000 times
Voltage AC 220V 、50/60Hz
Volume、weight 610x300x450mm、22KG
Imported high speed CCD 120 frm/s
Special technologies Sheath flow technology;Multithreading and edge recognition technology
Particle recognition speed ≥10000 particles /min
Analytical items (1) particle size distribution; typical data;maximum particle size;specific interval content;the particle size content more than or less than certain particle size.
(2) particle shape: aspect ratio and aspect ratio distribution; circularity and circularity distribution; particle image.
(3) particle count:To measure the number of particles in a certain volume of liquid.
Export items including the original parameter (including the sample information, testing information,etc),
analytical data (including the particle size distribution, aspect ratio, circularity and typical data,etc.),
distribution graph ( interval distribution histogram and cumulative distribution curve,etc.)