Adiabatic Dewar Calorimeter ADC II
หมวดหมู่สินค้า: Chilworth
รหัส : CTL02
31 กรกฎาคม 2565
ผู้ชม 203 ผู้ชม
Specification
Functional Specification and Deliverables:
- 6 kW 3-phase 380 V – 415 V Dewar oven. 0 – 400 °C capability
- Pressure range 0 – 100 barg. Vent relief set to 23 barg
- Dewar flask set with two fully ported and populated stainless steel head assemblies (ports include potential for entry of sample thermocouple, sample pressure sensor, sample heater, stirrer motor, additional spare thermocouple)
- Electronic control unit (configured to suit various test scenarios in conjunction with the supplied software) – with built in oven temperature controller (oven temperature tracks the sample temperature reliably within 1 °C or better under steady rate of rise conditions and can follow exothermic activity of sample at rates of up to 40 – 60 °C / min). Also features sample pressure and heater power local monitors, emergency manual vent for the vessel and interfacing to all ancillary Dewar services e.g system sensors and optional extras highlighted below
- Personal computer and LCD monitor with ADC software pre-installed
- Data acquisition card capable of monitoring and / or controlling: sample temperature, oven temperature, sample pressure, stirrer speed, stirrer torque, heater power and also optional extras such as Tempering cell pressure, tempering cell inlet and outlet solenoid operation, additional thermo-couples, Chemical injection pump, laboratory balance, Gas Burette. Typical log rates are: every 20 seconds for stan dard logging and every 0.1 second logging per channel based on the trigger point being exceeded on rate of pressure or temperature rise data
- Accuracy of pressure + / – 0.1 barg to 50 barg and + / – 0.2 barg above 50 barg
- Accuracy of temperature + / -1 °C
- 1 x instruction manual
More.