Applied Scientific Instruments Co., Ltd.

บริษัท แอพพลายด์ ไซแอนทิฟิค อินสตรูเม้นท์ส จํากัด

ศูนย์รวมเครื่องมือทดสอบ เครื่องมือวิทยาศาสตร์ เครื่องมือวิเคราะห์ทุกชนิด

 

ติดต่อสอบถาม

ผ่าน LINE ID

@asithailand

 

เครื่องมือวิเคราะห์ค่าความเข้มข้นปริมาณกำมะถันและไนโตรเจนรวม Next-Gen Nitrogen and Sulfur Analysis

หมวดหมู่สินค้า: Trace Elemental Instruments
รหัส : Xplorer-V-NS
The vertical combustion analyzer nitrogen sulfur offers the most features out of the smallest footprint. Completely optimized for your (liquid) hydrocarbon sample analysis. It offers fast combined TN/TS analysis, top detection performances, smart operation features, data-driven robustness, and a fully loaded method library. All in one single configuration. The analyzer produces accurate results rapidly and suggests user actions based on extensive data collection. A solution you can rely on. The Xplorer-V NS is designed to provide the best analytical experience when running (liquid) hydrocarbon applications. The analysis are conducted according to relevant international test methods, such

01 สิงหาคม 2565

ผู้ชม 1554 ผู้ชม

Summary.

Next-Gen Nitrogen and Sulfur Analysis

Xplorer-V NS; Vertical combustion analyzer nitrogen sulfur

The vertical combustion analyzer nitrogen sulfur offers the most features out of the smallest footprint. Completely optimized for your (liquid) hydrocarbon sample analysis. It offers fast combined TN/TS analysis, top detection performances, smart operation features, data-driven robustness, and a fully loaded method library. All in one single configuration.

The analyzer produces accurate results rapidly and suggests user actions based on extensive data collection. A solution you can rely on. The Xplorer-V NS is designed to provide the best analytical experience when running (liquid) hydrocarbon applications. The analysis are conducted according to relevant international test methods, such as ASTM, DIN, ISO, and UOP.

Benefits of the vertical combustion analyzer nitrogen sulfur.

  • Most Compact Analyzer on the Market
  • Integrated 26 Position Autosampler XLS-26
  • Fast Analysis: 2 – 6 Minutes
  • Xpro-V™ Combustion Tube
  • Ultra-Low Detection Limits: 10 ppb (S), 10 ppb (N)
  • Nitrogen Oxide Correction Technology (NO-CT)
  • Auto-Clean Function
  • Easily Accessible Filters and Permeation Tube
  • Data-Driven Robustness with TEIS Software Counter-Center™
  • Instrument Performance Dashboards
  • Operated through PC or Incorporated Touch Screen

Application.

  • Chemicals (e.g. Plastics, Aromatics, Olefins and Parafines)
  • Refinery Products (e.g. Crude Oil, Naphta, Lubricants, Gasoline, Diesel Fuel, Fuel Oil)
  • LPG and Gases (e.g. Pentane, Butane, Propane, Propylene)

Most Features out of the smallest Footprint
      The Xplorer-V utilizes a 26 position liquid autosampler, vertically oriented furnace, and best-in-class detectors to deliver spot-on liquid sample results in 2.5 to 6 minutes, depending on the application. Optionally, the analyzer can be configured with sampling systems for the automated analysis of LPG and Gaseous matrices. The various system components are perfectly aligned, enabling accurate sample introduction, combustion, and conditioning, which create the ideal circumstances for unparalleled detection performances. This powerful configuration comes together in the smallest combined Nitrogen and Sulfur analyzer on the market.

       Test Method Compliance & Typical Applications
       The demand for accurate analysis of low Nitrogen and Sulfur concentrations is growing. The Xplorer-V TN/TS easily meets this demand by producing unmatched results in repeatability- as well as reproducibility tests of different test methods and related matrices.
       Test Method Compliance at a Fingertip.
       TEIS Software features an extensive method library - system settings are automatically adjusted based on the selected method, following specific application- or international test method requirements.
       Nitrogen Oxide Elimination Technology (NO-ET™)
       NO-ET™ technology completely eliminates the possible interference of Nitrogen to the Sulfur signal as stipulated
in the ASTM D5453. A software wizard to calculate and correct for the possible Nitrogen interference is standard included in the Xplorer-V TN/TS.

 

Test Method Compliance.

Total Sulfur by Pulsed UVFluorescence : ASTM D5453,ASTM D6667,ASTM D7183,ASTM D7551,EN ISO 20846,NEN 17178,UOP 987-A.
Total Nitrogen by Chemiluminescence :
ASTM D4629, ASTM D5176, ASTM D7184, UOP 981, UOP 936


Specification.

Configurations • Xplorer-V TN/TS, Nitrogen and Sulfur analyzer
• Xplorer-V TN, Nitrogen analyzer
• Xplorer-V TS, Sulfur analyzer
Dimensions (W x H x D) 25 x 58 x 66 cm (9.8 x 23 x 26 inch)
Weight 40 kg (88 lbs.)
Voltage 100 - 240 VAC, 50 - 60 Hz
Power Requirement 750 W
Inert Gases O2 > 99.6 % (2.6)
Argon 99.998 % (4.8)
Helium 99.998 % (4.8)
Furnace Dual Zone, Max. 1150 ℃, 2100 ℉
Sample Matrices Liquids, Gas & LPG
Sample volume Liquids 5 - 250 µL, Gas 10 mL sample loop, LPG 50 or 100 µL sample loop
Detector Nitrogen Chemiluminescence
Detector Sulfur Xenon Pulsed, UV-Fluorescence
Flow Control Mass Flow
Software TEIS Software

 

ขอใบเสนอราคา

ดาวน์โหลดโบว์ชัวร์

 

Engine by shopup.com