Water determination in gases and LPG
31 กรกฎาคม 2565
ผู้ชม 1129 ผู้ชม
Summary
The Aquamax KF PRO LPG fulfils the requirements of the standard ASTM D 7995 - 19: Standard Test Method for Total Water in Liquid Butane by Liquefied Gas Sampler and Coulometric Karl Fischer Titration.The Aquamax KF PRO LPG fulfils the requirements of the standard ASTM D 7995 - 19: Standard Test Method for Total Water in Liquid Butane by Liquefied Gas Sampler and Coulometric Karl Fischer Titration.
The Aquamax KF PRO LPG is designed for an easy and accurate determination of water in liquefied and gaseous samples such as LPG and LNG.
The new device combines coulometric Karl Fischer method with an unique gas evaporation and dosing procedure.
The Aquamax KF PRO LPG includes all features required for ppm level water in LPG and Gas including, the sulphur removal cartridge eliminating the side reactions caused by sulphides/H2S. Our sample loop principle allows you to fully automate the measurements, up to 125 per day!
All Aquamax KF PRO LPG parts are totally enclosed making this system completely safe and robust for use in the demanding petroleum industry.
The unique ECH sample loop allows you to use the instrument in your laboratory with full automation, as a portable/field use analyzer or can be integrated in to your process as an on-line system.
Advantages
- Sulphur removal cartridge eliminating the side reactions caused by sulphides/H2S
- No interference calculation required
- 250 measurements can be performed in 48 hours
- Totally automated process, no operator input required for the test
- Suitable to test all gas types without any calibration or adjustments
- No separate rinsing gas is required
- Rinsing process is fully automated
- No balance is required
- High sample throughput and long reagent life
- Compact device
Application
LPG, LNG
- propane, propene, butane, butene, butadiene
- ethylene oxide
- chlorinated hydrocarbons, e. g. methylene chloride, ethylene chloride, vinyl chloride
Analysis of refrigerants:
- halogenated hydrocarbons
Analysis of permanent gases:
- natural gas
- technical gases
- mixtures of test gases
Specification
Technical Points | Description |
Measurement method | Coulometric Karl Fischer titration |
Sample | Pressurized gas sample (LNG, LPG) |
Sample dosing | Pressurized bottle or directly from the gas line |
Pressure reducer | internal (with heating element) |
Sample loop | 300 mL (gas) |
Rinsing and dosing | 0 … 15 steps for each, adjustable |
Measuring range | 1 μg ... 100 % |
Resolution | 0.1 ppm |
Detection limit | 1 ppm |
Power supply | 230 V/50 Hz; 115 V/60 Hz |
Dimensions | 33 x 49 x 48 cm (W x D x H) |
Weight | 24.0 kg |
Device control | PC software (PC not included in the scope of delivery) |