Water determination in oil and fuel samples
31 กรกฎาคม 2565
ผู้ชม 197 ผู้ชม
Summary
The Aquamax KF PRO Oil is the perfect instrument to measure ppm water in oils and fuels without the worry of interference side reactions caused by additives or sulphur/mercaptans. The unique “closed loop” principle means no additional carrier gas is necessary. Directly injecting the sample in to the oven means no blank value is required, making the Aquamax KF PRO Oil a truly accurate, trace level water in petroleum products titrator.
Learn more about your oils by using the temperature ramping program. This unique ECH feature allows you to
see all various types of bonded water, making the Aquamax KF Pro Oil the perfect tool in the R & D, Refining, Used Oil analysis laboratories and lubricant blending plants. Crude oils are also perfectly analysed as we have the ability to display both free and bonded water.
All Aquamax KF PRO Oil parts are totally enclosed making this system completely safe and robust for use in the demanding petroleum industry. The ECH technique allows a very long reagent life, because its capacity can be used completely.
Advantages
- Closed loop principle does not allow methanol to evaporate from the KF solvent
- Reagent capacity is used completely
- Additive and Sulphur side reactions minimized
- By using of temperature programs it is possible to separate free and chemically bonded water
- Temperature ramping program allows you to distinguish between various types of bonded water
- No blank value meaning true ppm accuracy
- Aquamax KF PRO Oil can be used in the laboratory or used as part of a mobile lab when taking a measurement from the sample point is critical
- Compact and rugged device
Application
- Transformer oils
- Hydraulic oils
- Engine oils
- Gear oils
- Crude oil
- Fuel oil
- Gasoline
- Diesel
- Jet fuel
- Bio fuel
- Petroleum products
- Insulating oils
- Silicone oil
- Lubricating oils
- Biopetroleum
- Biological oils
Specification
Technical Points | Description |
Measurement method | Coulometric Karl Fischer titration |
Sample administration | Manually with syringe |
Sample amount | 0.01 ... 20 mL |
Heating temperature | 35 ... 250 °C, isothermal or with temperature program |
Blank value | 0 µg |
Measuring range | 0.0001 ... 100 % |
Power supply | 230 V/50 Hz; 115 V/60 Hz |
Dimensions | 33 x 39 x 48 cm (W x D x H) |
Weight | 17 kg |
Device control | PC software (PC not included in the scope of delivery) |