เครื่องวัดค่าการหักเหแสงของตัวอย่าง และวัดค่าความหวาน (Digital Refractometer)
09 พฤศจิกายน 2565
ผู้ชม 1387 ผู้ชม
Summary
Bellingham + Stanley Digital Refractometer RFM970-T with Peltier Temperature Control and Touchscreen
- Wide Range – 1.30 to 1.70 RI
- Up to 6 decimal places RI
- Wide Temperature Range
- Peltier Temperature Control
- 7” High definition touchscreen
- Temperature Stability Checks
RFID user clearance
The RFM900-T Series refractometers are high precision instruments for use within petrochemical, pharmaceutical, flavours and fragrance industries. The ergonomically designed cases deal with any spills by the sloping shape of the case or by the PEEK spill barrier that surrounds the shallow, easy-clean stainless steel prism dish. The prism dish is shaped to hold sufficient sample for measurement whilst regaining a large enough surface area for rapid temperature stabilisation to occur. A Kalrez® gasket compliant with FDA / USP Class VI has been chosen as the prism seal due to its traceability and its almost impermeable nature to most chemicals. The sample cover forms a seal to prevent evaporation, particularly important when working with volatile solutions such as aromatic oils, and is interchangeable so that a presser type can be fitted for more viscous samples.
Inherent to the design is the high definition, capacitive touchscreen display that not only facilitates operation in factory environments, even when operated whilst wearing gloves, but also serves as the centrepiece for the new software. Combined, the touchscreen and GUI help the operator quickly manoeuvre through the user and configuration menus. On-screen graphical prompts support simple operation such as method loading, calibration and routine maintenance.
RFID swipe technology comes as standard, providing clearance and a log of operator and configuration functions – especially useful when being used in FDA-regulated environments operating in accordance with 21 CFR Part 11, or where good practice is adopted, e.g. as part of HACCP.
Measurements with the RFM900-T are made through a familiar measuring mode or by way of a simple Methods system with one-touch configuration of scale, temperature and data storage, with up to 8000 results being saved alongside pre-set limits that can be used as an audit tool. Methods may be user-defined or selected from the on-board library, such as the PHR-MEAN method that allows a number of individual readings from a single batch of pharmaceutical product to be measured with the resultant mean being displayed and recorded in accordance.
The RFM970-T Refractometer is particularly suited for operation within pharmaceutical environments, used to measure the purity of incoming raw materials as well as for production quality control in accordance with pharmacopeia (BP/EP/USP/JP).
RFM970-T refractometers may also be fitted with an optional flow through cell so that it may be used as part of a pharmaceutical pilot or full scale manufacturing process. This is because wetted materials traceability is made easy due to the FDA/USP Class VI components used as standard for areas where the sample is in contact with the instrument or flowcell. Materials Certificates for both refractometer and flowcell wetted materials are available.
Application
Bellingham + Stanley's digital refractometers cover many different applications. Specification varies from customer to customer which is why a range of refractometers have been designed offering a wide measuring range up to 1.70 RI with touchscreen interface on the highest spec model. Whilst ranges and applications may differ, build quality remains consistently high amongst Bellingham + Stanley’s RFM range. The same high-quality touchscreen is implemented on all T-Series models giving a high-definition user interface at your fingertips. All instruments are built with flat, easy-clean prism dishes to hold your sample, with any spills being dealt with by the ergonomically designed casing which allows samples to harmlessly drip away.
Specification
Measurement Specifications | ||
---|---|---|
RFM960-T | RFM970-T | |
Scales Refractive Index Sugar (°Brix) User-defined |
1.30 – 1.70 0-100 100 |
1.30 – 1.70 0-100 100 |
Resolution Refractive Index Sugar (°Brix) |
0.0001 0.1 |
0.000001 (selectable up to 6 d.p.) 0.01/0.001 |
Accuracy Refractive Index Sugar (°Brix) |
±0.0001 ±0.1 |
±0.00002 ±0.02 |
Precision Refractive Index Sugar (°Brix) |
± 0.00005 ± 0.05 |
± 0.000005 (6 d.p.) ± 0.005 |
User Scale Library | 20+ pre-programmed scales including HFCS (3), sugar (4), sucrose SG (3), ASTM D 2140, PHR-MEAN, Urea and more; plus customer programmable user scales via PC. | |
Reading Time | Minimum 4 seconds (dependant on precision required) | |
Measuring Temperature Range | 0°C or 10°C below ambient whichever is greater to 80°C | |
Temperature Sensor Accuracy | ± 0.03°C | |
Sample Temperature Stability | ± 0.02°C | |
Temperature Compensation Sucrose (°Brix) AG Fluid User Defined |
5 – 80 °C 5 – 40 °C Simple coefficient (units/°C) or polynomial function |
|
Temperature Stability Checks | None/delay time/repeatability/ Smart (independently selectable by Method) | |
Sample Illumination | Light Emitting Diode 589nm (100,000+ hours) | |
Prism / Dish | Artificial Sapphire / Stainless Steel 316 / PEEK (spill barrier) | |
Prism Seal | Kalrez® (FDA/USP Class VI compliant) | |
Interfaces | 3 x USB (A), 1 x USB (B), 1 x Ethernet, RS232 via USB adaptor (optional accessory) | |
Power | Instrument: 24 V DC, ±5%, <2A External PSU: 100-240V, 50-60Hz (supplied) |
|
Humidity Range | <90% RH (non condensing) |