Flash point by pensky-martens closed cup tester - automated

หมวดหมู่สินค้า: Normalab
รหัส : NPM450

23 เมษายน 2565

ผู้ชม 189 ผู้ชม

Summary

Efficient and compact, the NPM 450 is a fully automated Pensky-Martens Closed Cup Flash Point Tester with a combination of latest technologies. Due to the unique design, the NPM 450 offers you the best performance thanks to its predictive heating regulation and the air turbine enabling fast sample cooling after the test.

Thought for the user, the NPM 450 includes a revolutionary stirring and dipping arm, with an electric and gas ignition. A single combined system is delivered in standard. The user can automatically switch from gas to electric igniter (regulated automatically).

Characteristics 

  • Test Temperature Range : Ambient to 410°C
  • Flash point detection by thermocouple
  •  Electric and Gas ignition
  •  Regulated electric ignitor
  •  Integrated barometric correction
  • One hand easy operation
  • 6,4 inch High Definition screen
  • Capacitive touch screen
  • Standard Methods: A, B, C
  • Others: Quick A, Bitumen, Unknown Product, B Modified, Ambient
  •  User programmable methods (fast heating, automatic pre heating …)
  • Fast cooling
  • Sample cup holder stand

Security

  • Over-temperature protection
  • Automatic fire extinguishing system (Nitrogen supply is necessary)

Features

An intuitive software, coupled with a high definition screen, offers you a friendly, user interface, able to follow in real time the information of your tests.
Delivered in standard with the methods A, B & C, the NPM 450 offers you the possibility to create more than 1000 methods with all parameters (pre-heating, stirring, dipping….). Equipped with a capacitive touch screen, the NPM 450 can be used with gloves.
To ensure security, an overheating safety along with fire detection can trigger an automatic N2 extinguisher system. NPM 450 offers a secure operation of the unit.

Delivered With 

  •  Cup and Lid
  •  Pt 100 probe
  • Detection Thermocouple Cable
  • Electric Ignitor

Recommendations

Propane/Butane pressure: mini 20-40 mBar max (for gas ignition)
Nitrogen : mini 4 Bar, maxi 12 Bar (for fire extinguisher)

Technical Data

Power supply: AC 230V or AC115 V – 50/60 Hz – 4A
Dimensions: (W) 260 x (D) 460 x (H) 185 mm
Weight: 18 kg

 

Specification

  • Small size
  • Capacitive touch screean (6,4 inch HD)
  • One hand easy operation
  • Sample cup, probe, thermocouple holder stand
  • Test temperature range: 0 to 410°C
  • Unique combine ignition system (electric and gas)
  • Standard methods: A, B, C
  • Integrated barometric correction
  • Flash point detection by thermocouple
  • Automatic gas flame relighting
  • User programmable methods
  • Heavy sample preheating and automatic strirring start
  • Fast heating and cooling